MON630 Infant Incubator

SPECIFICATION
POWER
Model B-800-1E…110 / 120 Vac, 50Hz / 60Hz
Model B-800-2E…220 / 240 Vac, 50Hz / 60Hz
Power Input…≤ 400VA
Refer to IEC60601-1, this equipment belongs to ClassI, Type BF. Its operation is continuous.
ALARM
Power Alarm
Power Failure…Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle.
Fan Alarm
Fan Failure…Activates if Fan motor malfunctions.
Sensor Alarm
Air Sensor Failure…Air Sensor malfunction, refer unit to service.
Deviation Alarm
High Air Temperature…Activated if displayed Air Temperature is 1.5above set point.
Low Air Temperature…Activated if displayed Air Temperature is 2.5below set point.
Over Temperature Alarm
Over Air Temperature…Activated when displayed Air Temperature was approach 38 in Air Mode.
System Alarm
System Failure…Activated if Temperature Control System malfunctions.

Category:

Description

SPECIFICATION
POWER
Model B-800-1E…110 / 120 Vac, 50Hz / 60Hz
Model B-800-2E…220 / 240 Vac, 50Hz / 60Hz
Power Input…≤ 400VA
Refer to IEC60601-1, this equipment belongs to ClassI, Type BF. Its operation is continuous.
ALARM
Power Alarm
Power Failure…Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle.
Fan Alarm
Fan Failure…Activates if Fan motor malfunctions.
Sensor Alarm
Air Sensor Failure…Air Sensor malfunction, refer unit to service.
Deviation Alarm
High Air Temperature…Activated if displayed Air Temperature is 1.5above set point.
Low Air Temperature…Activated if displayed Air Temperature is 2.5below set point.
Over Temperature Alarm
Over Air Temperature…Activated when displayed Air Temperature was approach 38 in Air Mode.
System Alarm
System Failure…Activated if Temperature Control System malfunctions.